Tensile failure strain testing equipment for plastics is essential for determining the elongation behavior and ultimate deformation capacity of polymer materials under uniaxial tensile loading. This article provides a comprehensive technical overview of the instrumentation, specimen types, testing procedures, international standards, and industrial applications associated with measuring tensile strain at break in plastic materials.
Technical Overview
Tensile failure strain testing equipment for plastics refers to a class of laboratory instrumentation designed to apply a controlled uniaxial tensile load to a polymer specimen until fracture occurs, while simultaneously measuring the elongation of the specimen to determine the strain at break (tensile failure strain). The core instrument in this category is the universal testing machine (UTM), also known as a tensile tester or materials testing frame, which typically consists of the following key components:
- Load frame: A rigid, column-guided mechanical structure available in single-column (typically 1–5 kN capacity) or dual-column (5–600 kN capacity) configurations, selected based on the expected strength range of the plastic materials being tested.
- Load cell (force transducer): A precision strain-gauge-based sensor that measures the tensile force applied to the specimen. Load cells are available in multiple capacities and must comply with accuracy classes defined in ISO 7500-1 or ASTM E4.
- Crosshead drive system: A servo-motor-driven ball-screw mechanism that moves the crosshead at a precisely controlled, constant speed, typically ranging from 0.5 mm/min to 500 mm/min, in accordance with the requirements of ISO 527 or ASTM D638.
- Grips and fixtures: Mechanical wedge-action grips, pneumatic grips, or hydraulic clamps that securely hold the specimen ends without slippage or premature failure at the grip interface. Specialized fixtures are used for films, fibers, and elastomers.
- Extensometer: A device for directly measuring specimen elongation within a defined gauge length. Contact extensometers (clip-on types with knife edges) and non-contact video extensometers or laser extensometers are both widely used. Non-contact systems are particularly advantageous for measuring large strains at break in ductile plastics, as they avoid the risk of influencing specimen failure.
- Data acquisition and control system: Software that records force and displacement data at high sampling rates, calculates engineering stress and strain, constructs the stress-strain curve, and automatically determines tensile strength, strain at break, yield point, and tensile modulus.
Modern systems offer testing accuracy better than ±0.5% of the indicated force, strain resolution down to 1 µm, and fully automated test sequencing, including automatic specimen dimension measurement, statistical analysis of replicate specimens, and export of results in standardized report formats. Environmental chambers may be integrated to conduct tensile strain-at-break testing at elevated or sub-ambient temperatures, from -70 °C to +300 °C, which is critical for applications such as automotive and aerospace components.
Testing Samples
Molded Dumbbell Specimens
The most common specimen geometry for rigid and semi-rigid plastics is the dumbbell (dog-bone) shape, machined or injection-molded from plaques or directly from the molding process. ISO 527-2 defines specimen types 1A and 1B, while ASTM D638 defines Types I through V. Type 1A (ISO) and Type I (ASTM) specimens feature a gauge length of 50 mm (or 60 mm for ISO) and a preferred thickness of 4 mm.
Film and Sheet Specimens
Thin plastic films used in packaging are tested according to ISO 527-3 or ASTM D882, which prescribe rectangular strip specimens or specialized dumbbell shapes. These materials frequently exhibit very high strain-at-break values, often exceeding 300–600%, requiring long-travel extensometers or grip separation measurement with appropriate correction factors.
Injection-Molded and Machined Test Pieces
For quality control of molded parts, specimens may be cut directly from finished components. Multi-purpose test specimens produced in accordance with ISO 20753 or ASTM D638 ensure reproducibility between laboratories. Orientation effects from molding must be considered, as strain at break can differ significantly between machine-direction and transverse-direction specimens.
Specialty Specimens
Pipes are sampled as longitudinal strips per ISO 6259, geosynthetics and geomembranes are tested per ISO 10319, and reinforced plastics and composites follow ISO 527-4 and ISO 527-5, which use rectangular tabbed specimens with bonded end tabs to prevent grip-induced failure.
Testing Items
Tensile failure strain testing equipment is capable of determining a complete suite of mechanical properties from a single tensile test, including:
- Tensile strain at break (failure strain): The elongation of the gauge length at the moment of specimen rupture, expressed as a percentage of the original gauge length. This is the primary output parameter.
- Tensile strength: The maximum tensile stress sustained by the specimen before rupture.
- Tensile stress at yield: The stress at which the specimen continues to elongate without an increase in stress, characteristic of ductile thermoplastics.
- Tensile strain at yield: The strain corresponding to the yield point.
- Tensile modulus (Young's modulus): The slope of the initial linear portion of the stress-strain curve, determined between two specified strain limits, typically 0.0005 and 0.0025, using an extensometer.
- nominal strain at break: Strain calculated from crosshead displacement when extensometer measurement is impractical, subject to documented corrections.
- Poisson's ratio (optional): Measured with transverse extensometers or digital image correlation (DIC) systems.
Testing Methods
The internationally recognized test methods for determining tensile properties, including strain at break, of plastics are:
- ISO 527-1 / ISO 527-2: The primary international standard for tensile properties of plastics. ISO 527-1 specifies general principles, while ISO 527-2 covers test conditions for molding and extrusion plastics. The standard test speed is 1 mm/min for modulus determination and 50 mm/min (Type 1A/1B specimens) for strength and strain measurement.
- ASTM D638: The North American standard for tensile properties of unreinforced and reinforced plastics, specifying Type I through V specimens and standard crosshead speeds of 5, 50, or 500 mm/min depending on specimen type and material.
- GB/T 1040.1 / GB/T 1040.2: The Chinese national standard, technically aligned with ISO 527, widely used in domestic quality control and accreditation programs.
- ISO 527-3: For films and sheets under 1 mm thickness.
- ASTM D882: For tensile properties of thin plastic sheeting and films.
- JIS K 7161 / JIS K 7162: Japanese industrial standards harmonized with ISO 527-1 and ISO 527-2.
- ISO 6259: For tensile properties of thermoplastics pipes.
Typical test procedure: The specimen is conditioned at 23 °C ± 2 °C and 50% ± 10% relative humidity for a minimum of 24 hours (40 hours for ASTM). Specimen width and thickness are measured at three points within the gauge length to calculate the cross-sectional area. The specimen is mounted in the grips, the extensometer is attached to the gauge length, and the crosshead is set to the standard speed. The test proceeds until specimen rupture, during which force-extension data are continuously recorded. Strain at break is calculated as the extensometer-measured elongation at rupture divided by the initial gauge length, multiplied by 100%. A minimum of five valid specimens is tested, and the arithmetic mean, standard deviation, and coefficient of variation are reported. Specimens that fail outside the gauge length are typically rejected and retested.
Applications
Tensile failure strain data is fundamental to material selection, product design, and quality assurance across numerous industries:
- Automotive: Qualification of interior and exterior plastic components, bumpers, fuel system parts, and under-hood components where ductility at both ambient and low temperatures is a critical safety requirement.
- Packaging: Evaluation of polyethylene and polypropylene films, where elongation at break governs puncture resistance, formability, and seal performance in flexible packaging and stretch films.
- Medical devices: Characterization of catheter tubing, IV components, packaging blister films, and sterilization-resistant polymers in accordance with regulatory material qualification requirements.
- Electronics and electrical: Assessment of connector housings, cable insulation and jackets, and encapsulants, where strain at break indicates resistance to cracking during assembly and thermal cycling.
- Pipes and infrastructure: Long-term performance evaluation of PVC, PE, and PP piping systems, including testing of pipe walls and fusion joints.
- Construction and geosynthetics: Verification of geomembranes, geotextiles, and plastic building products against specified elongation criteria.
- Material development and research: Comparison of resin formulations, effects of additives, fillers, recycled content, and processing parameters on ductility; generation of input data for finite element analysis (FEA) and material cards.
- Quality control: Incoming resin inspection and production lot release testing performed by accredited third-party laboratories and in-house quality departments.
FAQ
What is tensile failure strain and why is it important?
Tensile failure strain (strain at break, or elongation at break) is the percentage of elongation a plastic specimen sustains at the moment it fractures under tensile loading. It is a direct measure of material ductility: high values indicate a tough, deformable material, while low values indicate brittleness. This parameter is essential for predicting how a plastic part will respond to impact, bending, thermal expansion mismatch, and assembly stresses.
How is strain at break measured accurately?
Accurate measurement requires an extensometer attached directly to the specimen gauge length, because grip displacement includes machine and grip compliance errors. For ductile plastics with very high elongation, non-contact video extensometers with automatic fiducial tracking are preferred, as contact devices may slip or interfere with the failure point. The strain at break is calculated from the extensometer reading at rupture divided by the initial gauge length.
What are the differences between ISO 527 and ASTM D638 testing?
Both standards measure the same fundamental properties, but they differ in specimen geometry (ISO 1A/1B versus ASTM Types I–V), gauge lengths (50 mm versus 60 mm for the standard specimens), testing speeds, and slightly different definitions of tensile modulus strain ranges. Because results are not directly interchangeable, the applicable standard must be specified when comparing data or submitting samples to an accredited third-party laboratory.
What crosshead speed should be used for tensile strain testing?
The standard speeds are defined by the governing test method. ISO 527-2 specifies 1 mm/min for modulus determination and 50 mm/min for determining strength and strain at break for standard specimens. ASTM D638 commonly specifies 5 mm/min or 50 mm/min depending on specimen type. Testing at non-standard speeds is permitted for research purposes, but the speed must be reported, as strain at break is strongly rate-dependent.
Why do strain-at-break results sometimes show large scatter?
High variability is typical because failure strain is highly sensitive to specimen preparation quality, surface defects, machining marks, internal voids, residual stresses from molding, moisture content (particularly with polyamides), and temperature. Testing a sufficient number of specimens (typically five or more), careful specimen preparation, conditioning per ISO 291 or ASTM D618, and proper grip alignment all reduce scatter.
What equipment capacity is needed for plastic tensile testing?
For most commodity and engineering thermoplastics tested with standard specimens, a 5 kN or 10 kN dual-column universal testing machine covers the range adequately. Highly filled or reinforced composites, or testing at extreme temperatures, may require higher capacity frames of 50 kN or more, while thin films can be tested on 500 N or 1 kN single-column systems. Load cell selection should ensure that the breaking force falls between 20% and 80% of the selected load cell range for optimal accuracy.
Can strain at break be tested at low or elevated temperatures?
Yes. Environmental chambers integrated with the universal testing machine allow testing from approximately -70 °C to +300 °C. This is particularly important for ductile-to-brittle transition assessment in automotive exterior parts, cold-climate applications, and heat-aging studies. Special high-temperature extensometers or non-contact optical systems are used for strain measurement inside the chamber.